HPLCColumnandSystemTroubleshooting
色譜柱和色譜系統的故障檢修WhatEveryHPLCUserShouldKnow每個HPLC使用者應該知道什么
2022/12/91HPLCColumnandSystemTroublHPLCComponents
HPLC的組成
Pump泵
Injector/Autosampler進樣器/自動進樣器
Column色譜柱
Detector檢測器
DataSystem/Integrator數據系統/積分儀表Allofthesecomponentscanhaveproblemsandrequiretroubleshooting.
所有這些部件可能出現故障而需要檢修。2022/12/92HPLCComponents
HPLC的組成PCategoriesofColumnProblems
色譜柱問題的類型
A.Pressure壓力
B.Peakshape峰形
C.Retention保留時間2022/12/93CategoriesofColumnProblemsPressureIssues
壓力問題ColumnObservationsPotentialProblems色譜柱觀測潛在的問題HighpressurePluggedfrit壓力高堵塞濾頭
Columncontamination
色譜柱污染
Piuggedpacking
堵塞填料2022/12/94PressureIssues
壓力問題ColumnObDeterminingtheCauseandCorrectingHighBackPressure查出原因糾正壓力Checkpressurewith/withoutcolumn-manypressureproblemsareduetoblockagesinthesystemorguard有/無柱子時檢查壓力-許多壓力問題是由于系統或保護柱的阻塞IfColumnPressureishigh如果柱壓高:Washcolumn-Eliminatecolumncontaminationand
沖洗柱子pluggedpacking排除柱污染和填料堵塞-highmolecularweight/adsorbedcompounds高分子量/被吸附化合物-precipitatefromsampleorbuffer
樣品或緩沖液中的沉淀物Backflushcolumn-Clearpluggedfrit反沖柱子-清潔阻塞的濾頭Changefrit-Clearpluggedfrit更換濾頭2022/12/95DeterminingtheCauseandCorrColumnCleaning柱子清洗Flushwithstrongersolventsthanyourmobilephase用比你的流動相更強的溶劑沖洗
Reversed-PhaseSolventChoicesinOrderofIncreasingStrength為了增加強度選擇反相溶劑Useatleast25mlofeachsolventforanalyticalcolumns對于分析柱每種溶劑至少用25ml沖洗Mobilephasewithoutbuffersalts沒有緩沖鹽的流動相100%Methanol甲醇100%Acetonitrile乙腈75%Acetonitrile:25%Isopropanol75乙腈:25%異丙醇100%Isopropanol異丙醇100%MethyleneChloride★二氯甲烷100%Hexane★
已烷★
WhenusingeitherHexaneorMethyleneChloridethecolumnmustbeflushedwithIsopropanolbeforereturningtoyourresvered-phasemobilephase.當用已烷蔌二氯甲烷柱子時,必須先用異丙醇沖洗,然后再用你的反相流動相2022/12/96ColumnCleaning柱子清洗FlushwitColumnCleaning柱子清洗NormalPhaseSolventChoicesinOrderofIncreasingStrength為了增加強度選用正相溶劑Useatleast50mlofeachsolvent
每種溶劑至少用50ml
50%Methanol:50%Chloroform50%甲醇:50%氯仿100%EthylAcetate乙酸乙酯2022/12/97ColumnCleaning柱子清洗NormalPhasHowtoChangeaFrit
怎樣更換濾頭ColumnInletFrit柱進口濾頭ColumnBody柱身CompressionFerrule壓縮金屬套圈Weargloves戴手套Donotallowbedtodry不要讓底座干燥Donottouchthecolumn-bodyheatwillextrudepacking不要使柱身受熱,否則填料會噴出Donotovertighten不要旋得太緊FemaleEndFitting旋好尾端螺母MaleEndFitting旋好尾端螺頭2022/12/98HowtoChangeaFrit
怎樣更換濾頭CoPreventingBackPressureProblems
壓力問題的預防Usecolumnprotection柱保護
-Guardcolumns保護柱/預柱-In-linefilters在線過濾器
SamplePreparation樣品預處理Appropriatecolumnflushing
適當的柱沖洗Filterbufferedmobilephase
過濾緩沖流動相2022/12/99PreventingBackPressureProblPreventingBackPressureProblems:In-LineDevices
壓力問題的預防:在線裝置MobilePhasePre-ColumnInjectorFilterGuardFromPump預柱進樣器濾頭Column從泵來的流動相保護柱
AnalyticalColumnFilterandGuardColumnActonSample分析柱濾頭和保護柱作用于樣品
Pre-ColumnActsonMobilePhase
預柱作用于流動相ToDetector
到檢測器2022/12/910PreventingBackPressureProblPreventingBackPressureProblems:SamplePreparation
壓力問題的預防:樣品制備Solvent/ChemicalEnvironment
溶劑/化學環境Particulate/AggregateRemove微粒/凝聚物的除去Filtersamples過濾樣品Centrifugation離心分離SolidPhaseExtraction(S.P.E.)
固相萃取CartridgesorPlates薄膜或薄層板DisksorMembranes圓盤或生物膜2022/12/911PreventingBackPressureProblPeakShapeIssues
峰形問題
Splitpeaks裂峰
Peaktailing峰拖尾
Broadpeaks寬峰Poorefficiency低柱效Manypeakshapeissuesalsocombinations-I.e.Broadandtailingortailingwithincreasedretention許多峰形問題是結合在一起的-例如:展寬和拖尾或拖尾和保留時間增加2022/12/912PeakShapeIssues
峰形問題SSplitPeaks裂峰Canbecausedby:可能的原因:Columncontamination柱污染Partiallypluggedfrit部分阻塞濾片Columnvoid柱頭塌陷Injectionsolventeffects溶劑效應2022/12/913SplitPeaks裂峰CanbecausedbSplitPeaks裂峰
ColumnContamination柱污染Column:StableBondSB-C8,4.6×250mm,5μmMobilePhase:60%25mMNa2HPO4,pH3.0:40%MeOHFlowRate:1.0mL/minTemperature:35oCDetection:UV254nmSample:FilteredOTCColdMedication:1.Pseudoephedrine2.APAP3.Unknown4.ChlorpheniramineInjection1峰3峰形較好Injection30峰3裂峰Injection1AfterColumnWashwith100%CAN經100%乙腈沖洗后峰3峰形極好2022/12/914SplitPeaks裂峰
ColumnContamiSplitPeaks裂峰
InjectionSolventEffects溶劑化效應Column:StableBondSB-C8,4.6×250mm,5μmMobilePhase:82%H2O:18%ACNInjectionVolume:30μLSample:1.Caffeine咖啡因2.Salicylamide水楊酰胺A.InjectionSolventB.InjectionSolvent100%AcetonitrileMobilePhase峰形寬拖尾峰形尖銳對稱樣品溶劑盡可能與流動相匹配
2022/12/915SplitPeaks裂峰
InjectionSolvDeterminingtheCauseofSplitPeaks
裂峰原因的確定1.Complexsamplematrixormanysamplesanalyzed-likelycolumncontaminationorpartiallypluggedfrit復雜樣品的基質或分析許多樣品-柱污染或部分阻塞濾片
2.MobilephasepH>=7-likelycolumnvoidduetosilicadissolution(unlessspecialtycolumnused)流動相pH>=7-由于硅膠溶解使柱塌陷(除非使用專門的柱子)
3.Injectionsolventstrongerthanmobilephase-likelysplitandbroadpeaks,dependentonvolume溶劑比流動相的可能性大-裂峰和寬峰,由樣品量來決定
2022/12/916DeterminingtheCauseofSplitPeakTailing,BroadeningandLossofEfficiency
峰拖尾、變寬和柱效降低Canbecausedby:可能的原因
Column“secondaryinteractions”柱“次級效應”Columnvoid柱塌陷Columncontamination柱污染Columnaging柱老化Columnloading柱負荷超載Extra-columneffects柱外效應2022/12/917PeakTailing,BroadeningandLPeakTailingColumn“SecondaryInteractions”
峰拖尾柱“次級效應”Column:Alkyl-C8,4.6×150mm,5μmMobilePhase:85%25mMNa2HPO4,pH7.0:15%ACNFlowRate:1.0mL/minTemperature:35oCSample:
1.Phenylpropanolamine苯丙醇胺/去甲麻黃堿2.Ephedrine麻黃堿3.Amphetamine安非他明/苯丙胺4.Methamphetamine脫氧麻黃堿5.Phenteramine苯三胺NoTEA無三乙胺10mMTEA10mM三乙胺USPTF(5%)美國藥典拖尾因子USPTF(5%)美國藥典拖尾因子1.1.2.9.6.354.1.265.1.575.1.14Peaktailingeliminatedwithmobilephasemodifier(TEA)atpH7用流動相減尾劑(三乙胺)在pH7消除峰拖尾2022/12/918PeakTailingColumn“SecondaryPeakTailingColumn“SecondaryInteractions”
峰拖尾柱“次級效應”Column:Alkyl-C8,4.6×150mm,5μmMobilePhase:85%25mMNa2HPO4,pH7.0:15%ACNFlowRate:1.0mL/minTemperature:35oCSample:1.Phenylpropanolamine苯丙醇胺/去甲麻黃堿2.Ephedrine麻黃堿3.Amphetamine安非他明/苯丙胺4.Methamphetamine脫氧麻黃堿5.Phenteramine苯三胺
pH3.0pH7.0USPTF(5%)美國藥典拖尾因子USPTF(5%)美國藥典拖尾因子4.1.334.2.35ReducingthemobilephasepHreducesinteractionswithsilanolsthatcausepeaktailing降低流動相的pH,減小與硅醇基作用引起的峰拖尾2022/12/919PeakTailingColumn“SecondaryPeakTailing峰拖尾
ColumnContamination柱污染Column:StableBondSB-C8,4.6×250mm,5μmMobilePhase:20%H2O:80%MeOHFlowRate:1.0mL/minTemperature:R.T.Detection:UV254nmSample:1.Uracil尿嘧啶2.Phenol苯酚3.4-Chloronitrobenzene4-氯硝基苯4.Toluene甲苯PlatesTFPlatesTFPlatesTF理論板數拖尾因子理論板數拖尾因子理論板數拖尾因子1.76292.081.79061.431.74481.062.120431.642.124431.212.122371.213.137271.693.179991.193.153661.114.133551.324.170981.254.190671.17QCtestforwarddirectionQCtestreversedirectionQCtestaftercleaning100%IPA,35oC
向前方向相反方向
100%異丙醇沖洗后2022/12/920PeakTailing峰拖尾
ColumnContamPeakTailing/Broadening
SampleLoadEffects
峰拖尾/變寬樣品過載效應Column:4.6×150mm,5μmMobilePhase:40%25mMNa2HPO4,pH7.0:60%ACNFlowRate:1.5mL/minTemperature:40oCSample:1.Desipramine去郁敏/脫甲基丙米嗪2.Nortriptyline
3.Doxepin4.Imipramine丙米嗪5.Amitriptyline阿米替林6.TrimipramineTailingEclipseXDB-C8Broadening/CompetitiveC8USPTF拖尾因子Plates理論板數HighLoad/10LowLoadHighLoad/10LowLoad高載負/10倍低載負高載負/10倍低載負1.1.601.7085059412.2.001.9081578423.1.561.56277662302539835627351005518910725
2022/12/921PeakTailing/Broadening
PeakBroadening,Splitting
ColumnVoid
峰變寬,裂開
柱塌陷MobilePhase:流動相:50%CAN:50%Water:0.2%TEA(~pH11)50%乙腈:50%水:0。2%三乙胺(~pH11)
After30injectionsInitial開始時
進樣30次后Multiplepeakshapechangescanbecausedbythesamecolumnproblem.InthecaseavoidresultedfromsilicadissolvedathighpH.多數峰形改變可能是同一柱問題引起的。在這個例子中,高pH溶解硅膠導致塌陷2022/12/922PeakBroadening,SplittingColuBroadPeaksUnknown“Phantom”PeaksColumn:Extend-C18,4.6×150mm,5μmMobilePhase:40%10mMTEA,pH11:60%MeOHFlowRate:1.0mL/minTemperature:R.T.Detection:UV254Sample:1.Maleat順丁烯二酸2.Pseudeophedrine偽麻黃堿3.Chlorphenniramine氯非尼拉明Plates理論板數1.59222.98793.779“Phantom”peakfromfirstinjection前次進樣未出完的峰Theextremelylowplatesareanindicationofanextremelylateelutingpeakfromtheprecedingrun.極低的理論板數表明這是一個前面進樣極遲洗脫出來的峰2022/12/923BroadPeaksUnknown“Phantom”PeakTailing峰拖尾
InjectorSealFailure進樣器密封性差Column:Bonus-RP,4.6×75mm,3.5μmMobilePhase:30%H2O:70%MeOHFlowRate:1.0mL/minTemperature:R.T.Detection:UV254nmSample:1.Uracil尿嘧啶2.Phenol苯酚3.N,N-DimethylanilineN,N-二甲苯胺PlatesTFPlatesTF理論板數拖尾因子理論板數拖尾因子1.22351.721.36701.452.34911.482.104571.093.54321.153.100851.00BeforeAfterreplacingrotorsealandisolationseal之前轉動桿和隔離密封墊檢修后Overdueinstrumentmaintenancecancausepeakshapeproblems.儀器保養超期可能引起峰形問題/進樣器劃痕2022/12/924PeakTailing峰拖尾
InjectorSealPeakTailing峰拖尾
Extra-ColumnVolume柱外死體積效應Column:StableBondSB-C18,4.6×30mm,3.5μmMobilePhase:85%H2Owith0.1%TFA:15%CANFlowRate:1.0mL/minTemperature:35oCSample:1.Phenylalanine2.5-benzyl-3,6-dioxo-2-piperazine
3.Asp-phe4.Aspartame10μlextra-columnvolume50μlextra-columnvolume2022/12/925PeakTailing峰拖尾
Extra-ColumnDeterminingtheCauseofPeakTailing
峰拖尾原因的確定Evaluatemobilephaseeffects-altermobilephasepHandadditivestoeliminatesecondaryinteractions評價流動相效果-改變流動相pH和添加劑消除次級效應/流動相中組份與柱子相互作用Evaluatecolumnchoice-trycolumnwithhighpuritysilicaordifferentbondingtechnology評價柱選擇-試用高純度硅膠柱或不同鍵合技術柱Reducesampleload減小進樣量Eliminateextra-columneffects消除柱外效應Flushcolumnandcheckforaging/void沖洗柱子檢查柱子老化/塌陷2022/12/926DeterminingtheCauseofPeakRetentionIssues
保留時間問題Retentiontimechanges(tr)保留時間改變Capacityfactor(retention)changes(k′)
容量因子改變Selectivitychanges()選擇性改變
2022/12/927RetentionIssues
保留時間問題RetentiChangesinRetentionSameColumn,OverTime相同柱子上保留時間改變,超時Maybecauseby:可能的原因:Columnaging柱老化Columncontamination柱污染Insufficientequilibration不夠平衡Poorcolumn/mobilephasecombination柱/流動相組成差Changeinmobilephase改變流動相Changeinflowrate改變流速otherinstrumentissues其他儀器問題
2022/12/928ChangesinRetentionSameColuMobilephaseChangeCausesChangeinRetention
流動相改變引起保留時間改變60%MeOH:40%0.1%TFAFreshTFAAddedtoMobilePhaseVolatileTFAevaporated/degassedfrommobilephase.易揮發的三氟乙酸從流動相中蒸發/揮發。Replacingitsolvedproblem.換用新的流動相就解決問題。使用有機揮發性酸時,流動相應新配。2022/12/929MobilephaseChangeCausesChaColumnAging/EquilibrationCausesRetention/SelectivityChanges
柱老化/平衡不夠引起保留時間/選擇性改變Column1-Initial1號柱開始時Column1-NextDay1號柱第二天Column1-AfterCleaningwith1%H3PO41號柱用
1%H3PO4清洗
Theprimaryanalytewassensitivetomobilephaseagingofthecolumn.最初的分析對流動相老化柱子是靈敏的Thepeakshapewasasecondaryissueresolvedbyflushingthecolumn.沖洗柱子第二要解決的問題是峰形Retentionandpeakshapewereasexpectedaftercleaning.清洗后保留時間和峰形都達到了預期效果2022/12/930ColumnAging/EquilibrationCauDeterminingtheCauseofRetentionChangesonthesameColumn
同一柱上保留時間改變原因的確定1.DetermineK′,,andtrforsuspectpeaks對懷疑峰K′,和tr的確定2.Washcolumn沖洗柱子3.Testnewcolumn-notelotnumber試驗新柱-記錄一批數據4.Reviewcolumnequilibrationprocedures觀察柱平衡過程5.Makeupfreshmobilephase4andtest配制新鮮流動相再試驗6.Checkinstrumentperformance檢查儀器性能2022/12/931DeterminingtheCauseofRetenChangeinRetention/SelectivityColumn-toColumn
柱與柱之間保留時間/選擇性的改變Differentcolumnhistories(aging)
不同的柱歷史(老化)Insufficient/inconsistentequilibration平衡不夠/不一致Poorcolumn/mobilephasecombination
柱/流動相差Changeinmobilephase流動相改變Changeinflowrate流速改變Otherinstrumentissues其他儀器問題Slightchangesincolumnbedvolume(tronly)柱內體積輕微改變2022/12/932ChangeinRetention/SelectivitLot-to-LotSelectivityChange(pH)
批與批之間選擇性改變(pH)pH4.5-Lot1pH3.0-Lot1pH4.5-Lot2pH3.0-Lot2pH4.5showsselectivitychangefromlot-to-lotforbasiccompounds
pH4.5顯示批與批之間對堿性化合物的選擇性改變pH3.0showsnoselectivitychangefromlot-to-lot,indicatingsilanolsensitivityatpH4.5
pH3.0顯示批與批之間對堿性化合物的無選擇性改變,表明在pH4.5硅醇基的靈敏性2022/12/933Lot-to-LotSelectivityChangeHPLCColumnandSystemTroubleshooting
色譜柱和色譜系統的故障檢修WhatEveryHPLCUserShouldKnow每個HPLC使用者應該知道什么
2022/12/934HPLCColumnandSystemTroublHPLCComponents
HPLC的組成
Pump泵
Injector/Autosampler進樣器/自動進樣器
Column色譜柱
Detector檢測器
DataSystem/Integrator數據系統/積分儀表Allofthesecomponentscanhaveproblemsandrequiretroubleshooting.
所有這些部件可能出現故障而需要檢修。2022/12/935HPLCComponents
HPLC的組成PCategoriesofColumnProblems
色譜柱問題的類型
A.Pressure壓力
B.Peakshape峰形
C.Retention保留時間2022/12/936CategoriesofColumnProblemsPressureIssues
壓力問題ColumnObservationsPotentialProblems色譜柱觀測潛在的問題HighpressurePluggedfrit壓力高堵塞濾頭
Columncontamination
色譜柱污染
Piuggedpacking
堵塞填料2022/12/937PressureIssues
壓力問題ColumnObDeterminingtheCauseandCorrectingHighBackPressure查出原因糾正壓力Checkpressurewith/withoutcolumn-manypressureproblemsareduetoblockagesinthesystemorguard有/無柱子時檢查壓力-許多壓力問題是由于系統或保護柱的阻塞IfColumnPressureishigh如果柱壓高:Washcolumn-Eliminatecolumncontaminationand
沖洗柱子pluggedpacking排除柱污染和填料堵塞-highmolecularweight/adsorbedcompounds高分子量/被吸附化合物-precipitatefromsampleorbuffer
樣品或緩沖液中的沉淀物Backflushcolumn-Clearpluggedfrit反沖柱子-清潔阻塞的濾頭Changefrit-Clearpluggedfrit更換濾頭2022/12/938DeterminingtheCauseandCorrColumnCleaning柱子清洗Flushwithstrongersolventsthanyourmobilephase用比你的流動相更強的溶劑沖洗
Reversed-PhaseSolventChoicesinOrderofIncreasingStrength為了增加強度選擇反相溶劑Useatleast25mlofeachsolventforanalyticalcolumns對于分析柱每種溶劑至少用25ml沖洗Mobilephasewithoutbuffersalts沒有緩沖鹽的流動相100%Methanol甲醇100%Acetonitrile乙腈75%Acetonitrile:25%Isopropanol75乙腈:25%異丙醇100%Isopropanol異丙醇100%MethyleneChloride★二氯甲烷100%Hexane★
已烷★
WhenusingeitherHexaneorMethyleneChloridethecolumnmustbeflushedwithIsopropanolbeforereturningtoyourresvered-phasemobilephase.當用已烷蔌二氯甲烷柱子時,必須先用異丙醇沖洗,然后再用你的反相流動相2022/12/939ColumnCleaning柱子清洗FlushwitColumnCleaning柱子清洗NormalPhaseSolventChoicesinOrderofIncreasingStrength為了增加強度選用正相溶劑Useatleast50mlofeachsolvent
每種溶劑至少用50ml
50%Methanol:50%Chloroform50%甲醇:50%氯仿100%EthylAcetate乙酸乙酯2022/12/940ColumnCleaning柱子清洗NormalPhasHowtoChangeaFrit
怎樣更換濾頭ColumnInletFrit柱進口濾頭ColumnBody柱身CompressionFerrule壓縮金屬套圈Weargloves戴手套Donotallowbedtodry不要讓底座干燥Donottouchthecolumn-bodyheatwillextrudepacking不要使柱身受熱,否則填料會噴出Donotovertighten不要旋得太緊FemaleEndFitting旋好尾端螺母MaleEndFitting旋好尾端螺頭2022/12/941HowtoChangeaFrit
怎樣更換濾頭CoPreventingBackPressureProblems
壓力問題的預防Usecolumnprotection柱保護
-Guardcolumns保護柱/預柱-In-linefilters在線過濾器
SamplePreparation樣品預處理Appropriatecolumnflushing
適當的柱沖洗Filterbufferedmobilephase
過濾緩沖流動相2022/12/942PreventingBackPressureProblPreventingBackPressureProblems:In-LineDevices
壓力問題的預防:在線裝置MobilePhasePre-ColumnInjectorFilterGuardFromPump預柱進樣器濾頭Column從泵來的流動相保護柱
AnalyticalColumnFilterandGuardColumnActonSample分析柱濾頭和保護柱作用于樣品
Pre-ColumnActsonMobilePhase
預柱作用于流動相ToDetector
到檢測器2022/12/943PreventingBackPressureProblPreventingBackPressureProblems:SamplePreparation
壓力問題的預防:樣品制備Solvent/ChemicalEnvironment
溶劑/化學環境Particulate/AggregateRemove微粒/凝聚物的除去Filtersamples過濾樣品Centrifugation離心分離SolidPhaseExtraction(S.P.E.)
固相萃取CartridgesorPlates薄膜或薄層板DisksorMembranes圓盤或生物膜2022/12/944PreventingBackPressureProblPeakShapeIssues
峰形問題
Splitpeaks裂峰
Peaktailing峰拖尾
Broadpeaks寬峰Poorefficiency低柱效Manypeakshapeissuesalsocombinations-I.e.Broadandtailingortailingwithincreasedretention許多峰形問題是結合在一起的-例如:展寬和拖尾或拖尾和保留時間增加2022/12/945PeakShapeIssues
峰形問題SSplitPeaks裂峰Canbecausedby:可能的原因:Columncontamination柱污染Partiallypluggedfrit部分阻塞濾片Columnvoid柱頭塌陷Injectionsolventeffects溶劑效應2022/12/946SplitPeaks裂峰CanbecausedbSplitPeaks裂峰
ColumnContamination柱污染Column:StableBondSB-C8,4.6×250mm,5μmMobilePhase:60%25mMNa2HPO4,pH3.0:40%MeOHFlowRate:1.0mL/minTemperature:35oCDetection:UV254nmSample:FilteredOTCColdMedication:1.Pseudoephedrine2.APAP3.Unknown4.ChlorpheniramineInjection1峰3峰形較好Injection30峰3裂峰Injection1AfterColumnWashwith100%CAN經100%乙腈沖洗后峰3峰形極好2022/12/947SplitPeaks裂峰
ColumnContamiSplitPeaks裂峰
InjectionSolventEffects溶劑化效應Column:StableBondSB-C8,4.6×250mm,5μmMobilePhase:82%H2O:18%ACNInjectionVolume:30μLSample:1.Caffeine咖啡因2.Salicylamide水楊酰胺A.InjectionSolventB.InjectionSolvent100%AcetonitrileMobilePhase峰形寬拖尾峰形尖銳對稱樣品溶劑盡可能與流動相匹配
2022/12/948SplitPeaks裂峰
InjectionSolvDeterminingtheCauseofSplitPeaks
裂峰原因的確定1.Complexsamplematrixormanysamplesanalyzed-likelycolumncontaminationorpartiallypluggedfrit復雜樣品的基質或分析許多樣品-柱污染或部分阻塞濾片
2.MobilephasepH>=7-likelycolumnvoidduetosilicadissolution(unlessspecialtycolumnused)流動相pH>=7-由于硅膠溶解使柱塌陷(除非使用專門的柱子)
3.Injectionsolventstrongerthanmobilephase-likelysplitandbroadpeaks,dependentonvolume溶劑比流動相的可能性大-裂峰和寬峰,由樣品量來決定
2022/12/949DeterminingtheCauseofSplitPeakTailing,BroadeningandLossofEfficiency
峰拖尾、變寬和柱效降低Canbecausedby:可能的原因
Column“secondaryinteractions”柱“次級效應”Columnvoid柱塌陷Columncontamination柱污染Columnaging柱老化Columnloading柱負荷超載Extra-columneffects柱外效應2022/12/950PeakTailing,BroadeningandLPeakTailingColumn“SecondaryInteractions”
峰拖尾柱“次級效應”Column:Alkyl-C8,4.6×150mm,5μmMobilePhase:85%25mMNa2HPO4,pH7.0:15%ACNFlowRate:1.0mL/minTemperature:35oCSample:
1.Phenylpropanolamine苯丙醇胺/去甲麻黃堿2.Ephedrine麻黃堿3.Amphetamine安非他明/苯丙胺4.Methamphetamine脫氧麻黃堿5.Phenteramine苯三胺NoTEA無三乙胺10mMTEA10mM三乙胺USPTF(5%)美國藥典拖尾因子USPTF(5%)美國藥典拖尾因子1.1.2.9.6.354.1.265.1.575.1.14Peaktailingeliminatedwithmobilephasemodifier(TEA)atpH7用流動相減尾劑(三乙胺)在pH7消除峰拖尾2022/12/951PeakTailingColumn“SecondaryPeakTailingColumn“SecondaryInteractions”
峰拖尾柱“次級效應”Column:Alkyl-C8,4.6×150mm,5μmMobilePhase:85%25mMNa2HPO4,pH7.0:15%ACNFlowRate:1.0mL/minTemperature:35oCSample:1.Phenylpropanolamine苯丙醇胺/去甲麻黃堿2.Ephedrine麻黃堿3.Amphetamine安非他明/苯丙胺4.Methamphetamine脫氧麻黃堿5.Phenteramine苯三胺
pH3.0pH7.0USPTF(5%)美國藥典拖尾因子USPTF(5%)美國藥典拖尾因子4.1.334.2.35ReducingthemobilephasepHreducesinteractionswithsilanolsthatcausepeaktailing降低流動相的pH,減小與硅醇基作用引起的峰拖尾2022/12/952PeakTailingColumn“SecondaryPeakTailing峰拖尾
ColumnContamination柱污染Column:StableBondSB-C8,4.6×250mm,5μmMobilePhase:20%H2O:80%MeOHFlowRate:1.0mL/minTemperature:R.T.Detection:UV254nmSample:1.Uracil尿嘧啶2.Phenol苯酚3.4-Chloronitrobenzene4-氯硝基苯4.Toluene甲苯PlatesTFPlatesTFPlatesTF理論板數拖尾因子理論板數拖尾因子理論板數拖尾因子1.76292.081.79061.431.74481.062.120431.642.124431.212.122371.213.137271.693.179991.193.153661.114.133551.324.170981.254.190671.17QCtestforwarddirectionQCtestreversedirectionQCtestaftercleaning100%IPA,35oC
向前方向相反方向
100%異丙醇沖洗后2022/12/953PeakTailing峰拖尾
ColumnContamPeakTailing/Broadening
SampleLoadEffects
峰拖尾/變寬樣品過載效應Column:4.6×150mm,5μmMobilePhase:40%25mMNa2HPO4,pH7.0:60%ACNFlowRate:1.5mL/minTemperature:40oCSample:1.Desipramine去郁敏/脫甲基丙米嗪2.Nortriptyline
3.Doxepin4.Imipramine丙米嗪5.Amitriptyline阿米替林6.TrimipramineTailingEclipseXDB-C8
溫馨提示
- 1. 本站所有資源如無特殊說明,都需要本地電腦安裝OFFICE2007和PDF閱讀器。圖紙軟件為CAD,CAXA,PROE,UG,SolidWorks等.壓縮文件請下載最新的WinRAR軟件解壓。
- 2. 本站的文檔不包含任何第三方提供的附件圖紙等,如果需要附件,請聯系上傳者。文件的所有權益歸上傳用戶所有。
- 3. 本站RAR壓縮包中若帶圖紙,網頁內容里面會有圖紙預覽,若沒有圖紙預覽就沒有圖紙。
- 4. 未經權益所有人同意不得將文件中的內容挪作商業或盈利用途。
- 5. 人人文庫網僅提供信息存儲空間,僅對用戶上傳內容的表現方式做保護處理,對用戶上傳分享的文檔內容本身不做任何修改或編輯,并不能對任何下載內容負責。
- 6. 下載文件中如有侵權或不適當內容,請與我們聯系,我們立即糾正。
- 7. 本站不保證下載資源的準確性、安全性和完整性, 同時也不承擔用戶因使用這些下載資源對自己和他人造成任何形式的傷害或損失。
最新文檔
- 畜牧飼料企業服務體系建設與優化考核試卷
- 磷肥產品標準與檢測方法考考核試卷
- 紡織原料的綠色采購與可持續利用考核試卷
- 干部休養所服務質量管理考核試卷
- 天津現代職業技術學院《鋼琴基礎(1)》2023-2024學年第一學期期末試卷
- 上海農林職業技術學院《粵劇藝術賞析》2023-2024學年第二學期期末試卷
- 酒泉職業技術學院《馬克思主義與社會方法論》2023-2024學年第二學期期末試卷
- 南充科技職業學院《西班牙語精讀五》2023-2024學年第一學期期末試卷
- 山西老區職業技術學院《生物醫學傳感檢測系統設計》2023-2024學年第二學期期末試卷
- 新野縣2025年數學三下期末質量檢測試題含解析
- 廣東省2024-2025學年佛山市普通高中教學質量檢測物理試卷及答案(二)高三試卷(佛山二模)
- 【9數一模】2025年安徽合肥市第四十五中學九年級中考一模數學試卷(含答案)
- 2024年安徽馬鞍山技師學院專任教師招聘真題
- 電網工程設備材料信息參考價(2024年第四季度)
- 2024年浙江省中考社會試卷真題(含標準答案及評分標準)
- 2023年(第九屆)全國大學生統計建模大賽 論文模板及說明
- 棗樹桃小食心蟲
- 防蠅防鼠防蟲情況記錄表
- 大客戶營銷技巧ppt課件
- C++優秀課件PPT
- 團險新產品契約及核保細則
評論
0/150
提交評論